High-Temperature Furnaces with Molybdenum Disilicide Heating Elements and Fiber Insulation up to 1800 °C
High-Temperature Furnaces with Molybdenum Disilicide Heating Elements and Fiber Insulation up to 1800 °C
High-Temperature Furnaces with Molybdenum Disilicide Heating Elements and Fiber Insulation up to 1800 °C
High-Temperature Furnaces with Molybdenum Disilicide Heating Elements and Fiber Insulation up to 1800 °C

Share:


High-Temperature Furnaces with Molybdenum Disilicide Heating Elements and Fiber Insulation up to 1800 °C

Designed as tabletop models, these compact high-temperature furnaces have a variety of advantages. The first-class workmanship using high-quality materials, combined with ease of operation, make these furnaces all-rounders in research, for example, to melt small glass samples. 

  • Tmax 1600 °C, 1750 °C, or 1800 °C
  • Recommended working temperature 1750 °C (for models LHT ../18), increased wear and tear must be expected in case of working at higher temperatures
  • Dual shell housing made of textured stainless steel sheets with additional fan cooling for low surface temperature
  • High-quality molybdenum disilicide heating elements
  • Exclusive use of insulation materials without categorization according to EC Regulation No 1272/2008 (CLP). This explicitly means that alumino silicate wool, also known as “refractory ceramic fiber” (RCF), which is classified and possibly carcinogenic, is not used.
  • Adjustable air inlet opening, exhaust air opening in the roof
  • Type B thermocouple
  • Defined application within the constraints of the operating instructions
  • NTLog Basic for Nabertherm controller: recording of process data with USB-flash drive
  • Controller P570 (50 programs with each 40 segments)
Additional Equipment
  • Over-temperature limiter with adjustable cutout temperature as temperature limiter to protect the furnace and load
  • Protective gas connection to purge with non-flammable protective or reaction gases, not gas tight
  • Manual or automatic gas supply system
  • Process control and documentation via VCD software package for monitoring, documentation and control
Model Tmax Inner dimensions in mm Volume Outer dimensions1 in mm Connected Electrical Weight Heating time
  in °C w d h in l W D H2 load in kW connection in kg in min3
LHT 02/16 1600 90 150 150 2 470 630 760+260 3.0 1-phase 75 30
LHT 04/16 1600 150 150 150 4 470 630 760+260 5.2 3-phase4 85 25
LHT 08/16 1600 150 300 150 8 470 810 760+260 8.0 3-phase4 100 25
                         
LHT 01/17 D 1650 110 120 120 1 385 425 525+195 2.9 1-phase 28 35
LHT 03/17 D 1650 135 155 200 4 470 630 770+260 3.0 1-phase 75 30
                         
LHT 02/17 1750 90 150 150 2 470 630 760+260 3.0 1-phase 75 35
LHT 04/17 1750 150 150 150 4 470 630 760+260 5.2 3-phase4 85 30
LHT 08/17 1750 150 300 150 8 470 810 760+260 8.0 3-phase4 100 30
                         
LHT 02/18 1800 90 150 150 2 470 630 760+260 3.6 1-phase 75 60
LHT 04/18 1800 150 150 150 4 470 630 760+260 5.2 3-phase4 85 40
LHT 08/18 1800 150 300 150 8 470 810 760+260 9.0 3-phase4 100 40
1External dimensions vary when furnace is equipped with additional equipment. Dimensions on request.
2Including opened lift door4Heating only between two phases
3Heating time of the empty and closed furnace up to Tmax −100 K (connected to 230 V 1/N/PE rsp. 400 V 3/N/PE)


 

 Inquiry - High-Temperature Furnaces with Molybdenum Disilicide Heating Elements and Fiber Insulation up to 1800 °C